The charged particle identification obtained by the analysis of signals coming from the CsI(Tl) detectors of the CHIMERA 4π heavy-ion detector is presented. A simple double-gate integration method, with the use of the cyclotron radiofrequency as reference time, results in low thresholds for isotopic particle identification. The dependence of the identification quality on the gate generation timing is discussed. Isotopic identification of light ions up to Beryllium is clearly seen. For the first time also the identification of Z = 5 particles is observed. The identification of neutrons interacting with CsI(Tl) by (n, α) and (n, γ) reactions is also discussed.

Particle identification method in the CsI(Tl) scintillator used for the CHIMERA 4 pi detector

TRIFIRO', ANTONIO;TRIMARCHI, MARINA;
2002-01-01

Abstract

The charged particle identification obtained by the analysis of signals coming from the CsI(Tl) detectors of the CHIMERA 4π heavy-ion detector is presented. A simple double-gate integration method, with the use of the cyclotron radiofrequency as reference time, results in low thresholds for isotopic particle identification. The dependence of the identification quality on the gate generation timing is discussed. Isotopic identification of light ions up to Beryllium is clearly seen. For the first time also the identification of Z = 5 particles is observed. The identification of neutrons interacting with CsI(Tl) by (n, α) and (n, γ) reactions is also discussed.
2002
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11570/1608805
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 68
  • ???jsp.display-item.citation.isi??? 70
social impact