A 59-year-old woman with a history of congestive heart failure, who previously received a biventricular cardioverter–defibrillator was admitted with dyspnea and peripheral edema. She was noted on telemetry to have multiple self-terminated episodes of rapid ventricular pacing, consistent with pacemaker-mediated tachycardia. All episodes started after three consecutive ventricular extrasystoles. “Tracking preference” algorithm, designed to maintain atrial-tracked biventricular pacing in CRT devices, resulted turned on. It affected minimum PVARP and resulted the trigger for pacemaker mediated tachycardia. © 2016 Elsevier Inc. All rights reserved.

TRACKING PREFERENCE AS UNUSUAL TRIGGER OF PACEMAKER-MEDIATED TACHYCARDIA IN A RESYNCHRONIZATION DEVICE

Pasquale Crea
Investigation
;
Francesco Luzza
Supervision
;
2016-01-01

Abstract

A 59-year-old woman with a history of congestive heart failure, who previously received a biventricular cardioverter–defibrillator was admitted with dyspnea and peripheral edema. She was noted on telemetry to have multiple self-terminated episodes of rapid ventricular pacing, consistent with pacemaker-mediated tachycardia. All episodes started after three consecutive ventricular extrasystoles. “Tracking preference” algorithm, designed to maintain atrial-tracked biventricular pacing in CRT devices, resulted turned on. It affected minimum PVARP and resulted the trigger for pacemaker mediated tachycardia. © 2016 Elsevier Inc. All rights reserved.
2016
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11570/3140968
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 3
  • ???jsp.display-item.citation.isi??? ND
social impact