A 70-year-old man, who previously received a dual chamber pacemaker for paroxysmal AV block (Sorin Symphony DR 2550), was noted on telemetry to have multiple episodes of rapid ventricular pacing at approximately 120 bpm. Evaluation of the telemetry strips revealed that all of the rapid ventricular pacing episodes were initiated by brief runs of escape junctional rhythm. Programmed bradycardia parameters were AAI SafeR with lower rate limit of 50 bpm. © 2017 Elsevier Inc. All rights reserved.

SAFER AND ESCAPE JUNCTIONAL RHYTHM: A SINGULAR TRIGGER FOR PACEMAKER-MEDIATED TACHYCARDIA

Pasquale Crea
Investigation
;
Teresa Crea
Data Curation
;
Francesco Luzza
Supervision
;
2017-01-01

Abstract

A 70-year-old man, who previously received a dual chamber pacemaker for paroxysmal AV block (Sorin Symphony DR 2550), was noted on telemetry to have multiple episodes of rapid ventricular pacing at approximately 120 bpm. Evaluation of the telemetry strips revealed that all of the rapid ventricular pacing episodes were initiated by brief runs of escape junctional rhythm. Programmed bradycardia parameters were AAI SafeR with lower rate limit of 50 bpm. © 2017 Elsevier Inc. All rights reserved.
2017
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11570/3140975
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 3
  • ???jsp.display-item.citation.isi??? 3
social impact