The Directive 2013/35/EU establishes standards for workers exposed to static and time varying magnetic fields. These limits are based on ICNIRP guidelines expressed in terms of the electric field induced in the body. The complexity of this measurement led to theoretical models being developed. In this study, the experimental evaluation included varying magnetic field exposures for two classes of MRI workers. The measurements are conducted on four different MRI Systems including one 0.35 T, two 1.5 T, and one 3.0 T. Pocket magnetic dosimeters were used and it was carried out during routine conditions, emergency conditions, and cold-head maintenance/substitution. The acquired data has been processed and the corresponding dB/dt curves have been computed as the first time derivative of the dataset. The weighted peak approach was also implemented for the compliance assessment with regulatory limits. The dB/dt peak values have been compared with the reference level (RL) proposed by ICNIRP. The results show that the RL always exceeds during measurements on the 3.0 T scanner and sometimes on 1.5 T. In light of the foregoing, the diffusion of ultra-high field MRI scanners involves the introduction of behavioral rules that could be more useful than a numerical action level.

dB/dt Evaluation in MRI Sites: Is ICNIRP Threshold Limit (for Workers) Exceeded?

Acri, Giuseppe
Primo
Writing – Original Draft Preparation
;
Denaro, Lucia;Sansotta, Carlo;Ruello, Elisa;Salmeri, Francesca Maria;Vermiglio, Giuseppe;Testagrossa, Barbara
Ultimo
2018-01-01

Abstract

The Directive 2013/35/EU establishes standards for workers exposed to static and time varying magnetic fields. These limits are based on ICNIRP guidelines expressed in terms of the electric field induced in the body. The complexity of this measurement led to theoretical models being developed. In this study, the experimental evaluation included varying magnetic field exposures for two classes of MRI workers. The measurements are conducted on four different MRI Systems including one 0.35 T, two 1.5 T, and one 3.0 T. Pocket magnetic dosimeters were used and it was carried out during routine conditions, emergency conditions, and cold-head maintenance/substitution. The acquired data has been processed and the corresponding dB/dt curves have been computed as the first time derivative of the dataset. The weighted peak approach was also implemented for the compliance assessment with regulatory limits. The dB/dt peak values have been compared with the reference level (RL) proposed by ICNIRP. The results show that the RL always exceeds during measurements on the 3.0 T scanner and sometimes on 1.5 T. In light of the foregoing, the diffusion of ultra-high field MRI scanners involves the introduction of behavioral rules that could be more useful than a numerical action level.
2018
File in questo prodotto:
File Dimensione Formato  
dB_dt Evaluation in MRI Sites_Is ICNIRP Threshold.pdf

accesso aperto

Descrizione: Articolo principale
Tipologia: Versione Editoriale (PDF)
Licenza: Creative commons
Dimensione 1.38 MB
Formato Adobe PDF
1.38 MB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11570/3127335
Citazioni
  • ???jsp.display-item.citation.pmc??? 4
  • Scopus 13
  • ???jsp.display-item.citation.isi??? 10
social impact