An 80-year-old male, who previously received a dual chamber pacemaker, was referred to our ambulatory for dizziness and fatigue. The ECG obtained showed sinus rhythm, highly variable atrioventricular (AV) interval and alternation between spontaneous and paced ventricular complexes. A spike on the ascending part of the T wave was observed, suggesting ventricular undersensing. However, telemetry-supported pacemaker control showed inconstant atrial undersensing. (C) 2016 Elsevier Inc. All rights reserved.

SPIKE ON T WAVE! WHAT WENT WRONG?

Pasquale Crea
Investigation
;
Teresa Crea
Data Curation
;
Francesco Luzza
Supervision
;
2017-01-01

Abstract

An 80-year-old male, who previously received a dual chamber pacemaker, was referred to our ambulatory for dizziness and fatigue. The ECG obtained showed sinus rhythm, highly variable atrioventricular (AV) interval and alternation between spontaneous and paced ventricular complexes. A spike on the ascending part of the T wave was observed, suggesting ventricular undersensing. However, telemetry-supported pacemaker control showed inconstant atrial undersensing. (C) 2016 Elsevier Inc. All rights reserved.