Background: B-Laser™ is a novel atherectomy device that uses a solid-state third harmonic pulsed Nd:YAG laser with an output of 355 nm. Early data showed that the B-Laser™ is safe in treating a broad range of infrainguinal arterial lesions. We present the results of the EX-PAD-03 U.S. pivotal trial of the EXIMO B-Laser™. Methods: EX-PAD-03 is a prospective, single-arm, multi-center, international, open-label, clinical study. The study enrolled patients in the United States and Europe. The primary efficacy endpoint was the average reduction in residual diameter stenosis of >20% from baseline prior to any adjunctive therapy achieved by the B-Laser™ catheter alone. The primary safety endpoint was freedom from major adverse events (MAEs) defined as: unplanned target limb amputation above the ankle, clinically driven target lesion revascularization (CD-TLR) and cardiovascular related death. Results: A total of 97 subjects (107 lesions) were enrolled. Mean age was 70.5 years and 51% were males. Diabetes mellitus was present in 42.3%. Mean lesion length was 53.96 ± 43.18 mm and 26.2% had severe calcification. Lesions were de novo (79.4%), followed by in-stent restenosis (ISR) (15.9%) and non in-stent restenosis (4.7%). The mean percent stenosis at the target lesion as assessed by the Core lab was 85.7% ± 12.2 (femoro-popliteal 85.6 ± 12.8%; tibials 86.0 ± 9.6%). Post B-Laser™ and prior to adjunctive therapy, the mean percent stenosis at the target lesion was 52.1%. This resulted in a mean reduction from baseline to post B-Laser™ of 33.6% ± 14.2% meeting the primary efficacy endpoint goal. The freedom from MAE through the 30-day follow-up period after intervention was 98.9%. Per Core lab, there was no device-related distal embolization, dissections that required additional therapy, perforation, or pseudoaneurysm. Bailout stenting was 0.9%. A significant improvement from baseline in ABI (0.24 ± 0.18), Rutherford category (−1.79 ± 1.22) and WIQ (0.26 ± 0.28) were noted at 1 month. There was no target lesion revascularization and the patency was 96.8% by duplex ultrasound criteria at 30-day follow up. Conclusion: The Eximo B-Laser™ is effective and safe in ablating atherosclerotic and restenotic tissue for both above and below the knee obstructive arterial disease. The device has a high safety profile including a low risk of distal embolization.

Acute and 30-Day Safety and Effectiveness Evaluation of Eximo Medical's B-Laser™, a Novel Atherectomy Device, in Subjects Affected With Infrainguinal Peripheral Arterial Disease: Results of the EX-PAD-03 Trial

Micari A.;
2020-01-01

Abstract

Background: B-Laser™ is a novel atherectomy device that uses a solid-state third harmonic pulsed Nd:YAG laser with an output of 355 nm. Early data showed that the B-Laser™ is safe in treating a broad range of infrainguinal arterial lesions. We present the results of the EX-PAD-03 U.S. pivotal trial of the EXIMO B-Laser™. Methods: EX-PAD-03 is a prospective, single-arm, multi-center, international, open-label, clinical study. The study enrolled patients in the United States and Europe. The primary efficacy endpoint was the average reduction in residual diameter stenosis of >20% from baseline prior to any adjunctive therapy achieved by the B-Laser™ catheter alone. The primary safety endpoint was freedom from major adverse events (MAEs) defined as: unplanned target limb amputation above the ankle, clinically driven target lesion revascularization (CD-TLR) and cardiovascular related death. Results: A total of 97 subjects (107 lesions) were enrolled. Mean age was 70.5 years and 51% were males. Diabetes mellitus was present in 42.3%. Mean lesion length was 53.96 ± 43.18 mm and 26.2% had severe calcification. Lesions were de novo (79.4%), followed by in-stent restenosis (ISR) (15.9%) and non in-stent restenosis (4.7%). The mean percent stenosis at the target lesion as assessed by the Core lab was 85.7% ± 12.2 (femoro-popliteal 85.6 ± 12.8%; tibials 86.0 ± 9.6%). Post B-Laser™ and prior to adjunctive therapy, the mean percent stenosis at the target lesion was 52.1%. This resulted in a mean reduction from baseline to post B-Laser™ of 33.6% ± 14.2% meeting the primary efficacy endpoint goal. The freedom from MAE through the 30-day follow-up period after intervention was 98.9%. Per Core lab, there was no device-related distal embolization, dissections that required additional therapy, perforation, or pseudoaneurysm. Bailout stenting was 0.9%. A significant improvement from baseline in ABI (0.24 ± 0.18), Rutherford category (−1.79 ± 1.22) and WIQ (0.26 ± 0.28) were noted at 1 month. There was no target lesion revascularization and the patency was 96.8% by duplex ultrasound criteria at 30-day follow up. Conclusion: The Eximo B-Laser™ is effective and safe in ablating atherosclerotic and restenotic tissue for both above and below the knee obstructive arterial disease. The device has a high safety profile including a low risk of distal embolization.
2020
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11570/3179502
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 16
  • ???jsp.display-item.citation.isi??? 15
social impact