Background: Periodic quality control (QC) procedures are important in order to guarantee the image quality of radiological equipment and are also conducted using phantoms simulating human body. Objective: To perform (QC) measurements in intraoral imaging devices, a new and simple phantom was manufactured. Besides, to simplify QC procedures, computerized LabView-based software has been devised, enabling determination of image quantitative parameters in real time or during post processing. Material and Methods: In this experimental study, the novel developed phantom consists of a Polymethyl methacrylate (PMMA) circular insert. It is able to perform a complete QC image program of X-ray intraoral equipment and also causes the evaluation of image uniformity, high and low contrast spatial resolution, image linearity and artefacts, with only two exposures. Results: Three raters analyzed the images using the LabView dedicated software and determined the quantitative and qualitative parameters in an innovative and accu-rate way. Statistical analysis evaluated the reliability of this study. Good accuracy of the quantitative and qualitative measurements for the different intraoral systems was obtained and no statistical differences were found using the inter-rater analysis. Conclusion: The achieved results and the related statistical analysis showed the validity of this methodology, which could be proposed as an alternative to the com-monly adopted procedures, and suggested that the novel phantom, coupled with the LabView based software, could be considered as an effective tool to carry out a QC image program in a reproducible manner.

A novel phantom and a dedicated developed software for image quality controls in x-ray intraoral devices

Acri G.
Primo
;
Gurgone S.;Borzelli D.;Testagrossa B.
Ultimo
2021-01-01

Abstract

Background: Periodic quality control (QC) procedures are important in order to guarantee the image quality of radiological equipment and are also conducted using phantoms simulating human body. Objective: To perform (QC) measurements in intraoral imaging devices, a new and simple phantom was manufactured. Besides, to simplify QC procedures, computerized LabView-based software has been devised, enabling determination of image quantitative parameters in real time or during post processing. Material and Methods: In this experimental study, the novel developed phantom consists of a Polymethyl methacrylate (PMMA) circular insert. It is able to perform a complete QC image program of X-ray intraoral equipment and also causes the evaluation of image uniformity, high and low contrast spatial resolution, image linearity and artefacts, with only two exposures. Results: Three raters analyzed the images using the LabView dedicated software and determined the quantitative and qualitative parameters in an innovative and accu-rate way. Statistical analysis evaluated the reliability of this study. Good accuracy of the quantitative and qualitative measurements for the different intraoral systems was obtained and no statistical differences were found using the inter-rater analysis. Conclusion: The achieved results and the related statistical analysis showed the validity of this methodology, which could be proposed as an alternative to the com-monly adopted procedures, and suggested that the novel phantom, coupled with the LabView based software, could be considered as an effective tool to carry out a QC image program in a reproducible manner.
2021
File in questo prodotto:
File Dimensione Formato  
JBPE_Volume 11_Issue 2_Pages 151-162 (1).pdf

accesso aperto

Descrizione: Articolo principale
Tipologia: Versione Editoriale (PDF)
Licenza: Creative commons
Dimensione 1.24 MB
Formato Adobe PDF
1.24 MB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11570/3203729
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 2
  • ???jsp.display-item.citation.isi??? ND
social impact