The present research is specifically based on the use of greatly-reduced gas flows, in flow-modulator (FM) comprehensive two-dimensional gas chromatography systems. In particular, focus of the present research is directed to FM devices characterized by an accumulation stage, and a much briefer re-injection step. It has been widely accepted that the operation of such FM systems requires high gas flows (≥20. mL/min), to re-inject the gas-phase contents of sample (or accumulation) loops, onto the second column. On the contrary, it will be herein demonstrated that much lower gas flows (≈ 6-8. mL/min) can efficiently perform the modulation step of re-injection. The possibility of using such improved operational conditions is given simply by a fine optimization of the processes of accumulation and re-injection. The application of lower gas flows not only means that second-dimension separations are carried out under better analytical conditions, but, even more importantly, greatly reduces problems which arise when using mass spectrometry (i.e., sensitivity and instrumental pumping capacity).

Flow-modulation low-pressure comprehensive two-dimensional gas chromatography

TRANCHIDA, Peter Quinto;FRANCHINA, FLAVIO ANTONIO;DUGO, Paola;MONDELLO, Luigi
2014-01-01

Abstract

The present research is specifically based on the use of greatly-reduced gas flows, in flow-modulator (FM) comprehensive two-dimensional gas chromatography systems. In particular, focus of the present research is directed to FM devices characterized by an accumulation stage, and a much briefer re-injection step. It has been widely accepted that the operation of such FM systems requires high gas flows (≥20. mL/min), to re-inject the gas-phase contents of sample (or accumulation) loops, onto the second column. On the contrary, it will be herein demonstrated that much lower gas flows (≈ 6-8. mL/min) can efficiently perform the modulation step of re-injection. The possibility of using such improved operational conditions is given simply by a fine optimization of the processes of accumulation and re-injection. The application of lower gas flows not only means that second-dimension separations are carried out under better analytical conditions, but, even more importantly, greatly reduces problems which arise when using mass spectrometry (i.e., sensitivity and instrumental pumping capacity).
2014
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11570/3062235
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 42
  • ???jsp.display-item.citation.isi??? 36
social impact